Meeting 24-3-2016

Date: 24th feb 2016
Time: 15:00 UK time
Attendance: Adam, CHeidt, CHunt, Paul K, Melissa

Agenda and Minutes

Prev Actions:
DONE! Helium quantitative analysis and diffuser decision - MU
DONE! Cryo 3 - EO, AB, CMW
4 tested, no temp or bias calib as yet. Work to continue. Board testing in Tracker lab - EO

New Actions:
Look at first data - Melissa, Paul and Adam

1. Status of hardware,

Commissioning: EO CMW MU

  • cryostat status (& spares)
  • daresbury interlocks.
  • tracker daq/c&m systems

CRYO3 SWAP
Ed and Ao were doing a new calibration late last week, and spotted a particularly bad section (50 new bad channels) of cryostat 3, which has appeared since the warm up/cool down. Fortunately the channels are grouped closely together. Hence the idea to move the waveguides to a new connector, to avoid the bad channels on the board.

A waveguide swap to solve this problem happened this morning. Waveguide 29 was moved from the far right connector to the third from left. This relates to MCM8-MCM3, or in other words, WG 29 was connected to 5120-5183 and 6180-6143 and it has moved to 5440-5503 & 5760-5823. These channels are looking good so far and are currently being made light tight.

Mapping being changed by CHeidt and calib redone by Ed. Both will be checked with data and we should be ready to go by tomorrow.

This effects are in stn 1 plane 1

SPARE BOARD TEST STAND
Set up and being used
4 with good readout and firmware
Temp and bias calibration to come
Work will continue in around 1 month

MICE hall, inc Services and requirements : CM, GB,

  • Helium filling and diffuser
    Decision made, no diffuser changes
    He leak not above background level at TOF PMTs so no risk there.
    O2 in tracker DS measured before the bubbler to be 0.5% so that looks good (not too much air in the tracker).
    Need to fix hall probe leaks with silicon etc.
  • Tracker lab.

First Data Analysis - ALL
  • Feb data taking analysis so far
    Looks good from OnRec so far! DS tracker station 3 looks like its missing one side. DS S1 p1 is the problem. We will all look at data.4

2. Software: PK, AD, CHunt, CHeidt

  • Readiness
  • Validation script
  • CDB
  • Geometry,
  • Calibrations, Configurations
  • Monte Carlo
  • Pattern Recognition
  • Kalman
  • Online Displays
  • Performance and Analysis

Tracker Software Paper and Documentation - AD, PK DR

4. Alignment

Progress ongoing.

AoB

The ADC/Pedestal calibration summaries are available on: http://heplnv152.pp.rl.ac.uk/tracker/summary.xhtml
Follow the links for the more specific data. Some work is still ongoing to neaten up things.

New Actions

Phone Details

  • The four digit company id: 4673 (callers from outside the UK)
  • The pin number: 419357

Phone numbers:

UK (Imperial College --preferred) 0845 545 4460 English
Austria 0820 4000 1503 English
0820 4000 15 02 German
Belgium 070 35 99 45 English
070 35 99 94 Dutch
070 35 98 66 French
Canada 1 712 432 2832 English
1 712 432 2844 French
Germany 01803 001 178 English
01570 248 8303 English (mobile)
01803 001 177 German
01570 248 8101 German (mobile)
Ireland 0818 270 007 English
Italy 848 390 166 English
848 391 819 Italian
Netherlands 0870 001 901 English
0870 001 909 Dutch
Poland 0801 003 543 English
0801 003 533 Polish
South Africa 087 550 0375 English
Spain 902 885 318 English
902 881 200 Spanish
Sweden 0939 2066 400 English
0939 2066 300 Swedish
Switzerland 0848 560 190 English
0848 560 378 French
0848 560 347 German
0848 560 385 Italian
United Kingdom (PowWowNow) 0844 4 73 73 73 English
United States 1 415 363 0833 English

Updated by Overton, Edward almost 8 years ago ยท 3 revisions