20160720 settings

The spectrometer solenoids have been shown to be reasonably stable running at 140 A, with revised QPS trip levels. There is an aim to bring up the magnet line in some solenoid configuration, probably at 140 A. It is not clear how much time we will get running with the full line, how much we will be limited to only e.g. one module or e.g. only SSU and SSD. It is clear that we will not run match coils. So I have prepared a set of magnet/beam line settings for the eventuality that we have SSU and SSD both powered.

Consideration:

  • Match coils will not be available
  • End coils E1/E2 are expected to run at CC current
  • ECE is not allowed > 140 A
  • FC is not allowed > 90 A
  • It is likely that the magnets are not aligned to each other or detectors to better than ~ few mm
  • It is possible that there are current-dependent misalignment of the trackers due to forces on the SS

In this context, I have prepared some generic settings at fixed current where we do a momentum scan. This enables us to

  • Validate tracker reconstruction
  • Validate the detector alignment (TOF to tracker)
  • Validate the detector alignment (tracker to tracker, through the fields)

Then I have prepared some magnet scan settings. This enables us to

  • Validate the magnet alignment/model
  • Validate the detector alignment (TOF to tracker)

It will be challenging to untangle magnet movement with current from detector alignment. One could imagine doing the momentum scan at two different magnet current settings. For now, I think we should try these run settings and seek to figure out what is going on...

Task 1: Fixed magnet settings

Task Description Momentum # Triggers # shifts SSU (ECE) SSU M1/M2 FC SSD (ECE) SSD M1/M2 Run control tag
1a Magnet alignment - sol 200 200k at TOF2 0.5 ? N/N Solenoid ? N/N 3-200+M3-Test1
1b Magnet alignment - sol 140 200k at TOF2 0.5 ? N/N Solenoid ? N/N 3-140+M3-Test2
1c Magnet alignment - sol 240 200k at TOF2 0.5 ? N/N Solenoid ? N/N 3-240+M3-Test1
1d Magnet alignment - sol 300 200k at TOF2 0.2 ? N/N Solenoid ? N/N Custom
1e Magnet alignment - sol 400 200k at TOF2 0.03 ? N/N Solenoid ? N/N Custom

Current settings for Run 1d

Particle Species p at Tgt Proton Absorber Q1 Q2 Q3 D1 DS D2 Q4 Q5 Q6 Q7 Q8 Q9 Similar Run(s) Run Tag
MeV/c mm A A A A A A A A A A A A DAQ run #
Pion 300.0 82 85.6 107.1 74.8 261.1 445.8 131.5 244.3 300.6 199.4 204.7 309.9 264.9 8012, 8013, 8014 Individual settings

Current settings for Run 1e

Particle Species p at Tgt Proton Absorber Q1 Q2 Q3 D1 DS D2 Q4 Q5 Q6 Q7 Q8 Q9 Similar Run(s) Run Tag
MeV/c mm A A A A A A A A A A A A DAQ run #
Pion 400.0 82 113.7 142.2 98.9 370.1 743.66 170 213.1 285.7 189.5 193.7 293.3 250.6 8004 Individual settings

Task 2: Fixed magnet settings SSU and SSD

If SSU and SSD are powered and the magnet people agree; repeat the fixed magnet settings as in Task 1. This takes priority over task 3 below.

Task 3: Magnet Scans SSU and SSD

If SSU and SSD are powered and the magnet people agree. The idea is to change the magnet settings by +/- 5 % and +/- 10 % and watch how the beam responds.

Task Description Momentum # Triggers # shifts SSU (ECE) SSU M1/M2 FC SSD (ECE) SSD M1/M2 Run control tag
3a Magnet alignment - sol 200 100k at TOF2 0.5 120 A N/N N 130 A N/N 3-200+M3-Test1
3b Magnet alignment - sol 200 100k at TOF2 0.5 125 A N/N N 130 A N/N 3-200+M3-Test1
3c Magnet alignment - sol 200 100k at TOF2 0.5 135 A N/N N 130 A N/N 3-200+M3-Test1
3d Magnet alignment - sol 200 100k at TOF2 0.5 140 A N/N N 130 A N/N 3-200+M3-Test1
3e Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N N 125 A N/N 3-200+M3-Test1
3f Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N N 120 A N/N 3-200+M3-Test1
3g Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N N 135 A N/N 3-200+M3-Test1
3h Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N N 140 A N/N 3-200+M3-Test1

Task 4: Fixed magnet settings SSU, FC and SSD

If SSU, FC and SSD are powered and the magnet people agree; repeat the fixed magnet settings as in Task 1. Use ECE at 130 A and FC at 60.7 A; FC is set for reason of optical symmetry

Task 5: Diffuser settings

Aim is to validate the diffuser is working okay. For this setting, we are concerned to get a reasonable statistic at TKU; and watch to see the emittance change/energy loss when different absorber petals are installed. We only care about SSU operation for this setting (to try to measure change in pz).

We do a diffuser empty at the beginning and end to check for beamline stability. If this setting runs over several days, it is probably worth repeating the diffuser empty setting as a reference.

Task Description Diffuser Momentum # Triggers # shifts SSU (ECE) SSU M1/M2 FC SSD (ECE) SSD M1/M2 Run control tag
5a Diffuser 0000 200 100k at TOF1 0.2 140 N/N Solenoid ? N/N 3-200+M3-Test1
5b Diffuser 1000 200 100k at TOF1 0.2 140 N/N Solenoid ? N/N 3-200+M3-Test1
5c Diffuser 0100 200 100k at TOF1 0.2 140 N/N Solenoid ? N/N 3-200+M3-Test1
5d Diffuser 0010 200 100k at TOF1 0.2 140 N/N Solenoid ? N/N 3-200+M3-Test1
5e Diffuser 0001 200 100k at TOF1 0.2 140 N/N Solenoid ? N/N 3-200+M3-Test1
5f Diffuser 1111 200 100k at TOF1 0.2 140 N/N Solenoid ? N/N 3-200+M3-Test1
5g Diffuser 0000 200 100k at TOF1 0.2 140 N/N Solenoid ? N/N 3-200+M3-Test1

Task 6: Magnet Scans SSU, FC and SSD

If SSU, FC and SSD are powered and the magnet people agree. The idea is to change the magnet settings by +/- 5 % and +/- 10 % and watch how the beam responds.

Task Description Momentum # Triggers # shifts SSU (ECE) SSU M1/M2 FC SSD (ECE) SSD M1/M2 Run control tag
6a Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N 54.7 A 130 A N/N 3-200+M3-Test1
6b Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N 57.7 A 130 A N/N 3-200+M3-Test1
6c Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N 63.7 A 130 A N/N 3-200+M3-Test1
6d Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N 66.7 A 130 A N/N 3-200+M3-Test1
6e Magnet alignment - sol 200 100k at TOF2 0.5 120 A N/N 60.7 A 130 A N/N 3-200+M3-Test1
6f Magnet alignment - sol 200 100k at TOF2 0.5 125 A N/N 60.7 A 130 A N/N 3-200+M3-Test1
6g Magnet alignment - sol 200 100k at TOF2 0.5 135 A N/N 60.7 A 130 A N/N 3-200+M3-Test1
6h Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N 60.7 A 130 A N/N 3-200+M3-Test1
6i Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N 60.7 A 120 A N/N 3-200+M3-Test1
6j Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N 60.7 A 125 A N/N 3-200+M3-Test1
6k Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N 60.7 A 135 A N/N 3-200+M3-Test1
6l Magnet alignment - sol 200 100k at TOF2 0.5 130 A N/N 60.7 A 140 A N/N 3-200+M3-Test1

Updated by Franchini, Paolo over 6 years ago ยท 27 revisions