FF150613 TOF » History » Version 1

Rajaram, Durga, 16 June 2013 22:32

1 1 Rajaram, Durga
h1. FF150613 TOF
2
3
In Step I data, Ed sees that the e+ TOF(0->1) is 26ns instead of 25.6 (as predicted by post-Step I survey geometry)
4
Possible reasons are:
5
1) TOF z's used in Step I calibration do not match survey
6
2) the TDC->time conversion, which is known to be wrong by ~2 %
7
8
Solution:
9
Re-run calibration of Step I data (DR)
10
To do this we need to know 1) the calibration runs, and 2) the TOF positions in Step I
11
(Have asked Yordan and Mark which data-sets were used for the Step I calibration)
12
13
The calibration has to be re-run for the latest data, in any case.
14
15
The new calibrations should be uploaded to CDB (DR)
16
17
Implement getting calibrations by RUN (DR)
18
Implement getting a different calibration for the same RUN/time-range by looking at "id" (DR)