Feature #1701
Tracker mapping seems not to be complete?
Status:
Closed
Priority:
High
Assignee:
Category:
Tracker
Target version:
Start date:
30 June 2015
Due date:
% Done:
100%
Estimated time:
Workflow:
New Issue
Description
I have been running the tracker real data digitization, and added some cerr statements for when a DAQ hit is found, but matching mapping entry is available. There seem to be a good number for each trigger. Some example output from run 07002 using MAUS 0.9.6:
WARNING: UId 496 not present! chan_ro: 112, bank: 3, board: 0
WARNING: UId 504 not present! chan_ro: 120, bank: 3, board: 0
WARNING: UId 517 not present! chan_ro: 5, bank: 0, board: 1
WARNING: UId 994 not present! chan_ro: 98, bank: 3, board: 1
WARNING: UId 974 not present! chan_ro: 78, bank: 3, board: 1
etc etc.
Is this expected, or does it indicate a problem?
Updated by Dobbs, Adam about 8 years ago
- Status changed from Open to Closed
- % Done changed from 0 to 100
Solved, all is well, cause is unused electronics readout channels.